Page 18 MSI-4300 Porta-Weigh Plus/5 Key

User Guide

“BLANKLCD.” In this mode the liquid crystal
display weight reading blanks out and the message
displays “HI SETPT.” The MOSFET functions as
before. Use the SELECT key again to change to the
“BLINKLCD” response. Weight readings continue,
but the display alternates between a standard weight
reading and a “HI SET PT” indication. In all three
modes, the overload condition will override the
display mode.

8) After the desired setpoint response is selected, press

ENTER.

9) The Setpoint Menu is now back on the screen. The

SELECT key can be used to return to any of the
above operations (VALUE, MODE, RESPONSE)
if modifications need to be made. To quit and re-
turn to normal scale operation use SELECT until
the message “EXIT?” appears and push ENTER, or
push CLEAR (no shift required). This returns you
to the Setup Menu. Again, push CLEAR to get out
of the menu and back to standard scale operation.
Use the same procedure for a Low Setpoint.

Disable Setpoint:
It is often useful to be able to set up a Setpoint ahead of time and disable it for
testing purposes.

1) From the Setup function, push the SELECT (NET/

GROSS) key until the message reads “HI SETPT”
(or “LO SETPT”).

2) Push ENTER (TEST). This accepts that the HI

SETPT is to be modified.

3) Push SELECT until the “MODE” message appears.

Push ENTER.

4) Push the SELECT key until the message “DIS-

ABLE” appears. Push ENTER. This disables the
Setpoint and returns you to the Setpoint Menu.

5) Push CLEAR (ZERO) two more times to get out of

the Setpoint and Setup menus, or using the SE-
LECT key, scroll to the next desired operation. To
re-enable the Setpoint, follow steps 1-3 above. On
step 4 select the “GROSS” mode or the
“NETGROSS” mode.