Rice Lake Wheelchair Scale Technical Manual

8. Use the

BMI

key to move the flashing digit to the left or right.

9. Once all the digits have been entered, press the

ENTER

key and

Clear

is displayed.

10. Make sure the scale platform is clear of weight and press the

ENTER

key again and ===== is displayed.

11. A request to put the chosen load on the platform is displayed by

Put xxx.xx

.

12. Put the chosen weight on the platform and press the

ENTER

key. ===== is displayed and then

Save

.

13. Press the

ENTER

again and the display indicates

Done

.

14. Press the

BMI

key three times to exit back out to the top level

Done

parameter.

15. Press the

ENTER

key to return to weigh mode.

To exit calibration without changing zero or span existing calibration, press the

CLEAR

key, then the

BMI

key.

7.0

Test Menu

To access the

TEST

menu, use the following steps.

1. Turn the scale on by pressing the

On/Off

key and the

ENTER

key simultaneously until

ID

flashes.

2. Press the

ENTER

key again.

3. Continue to press the

BMI

key to scroll through the various menu items.

4. Once complete, press the

BMI

key again and

Done

is displayed.

5. Press the

ENTER

key to start the weighing process.

Figure 7-1. Test Menu

Test Menu

Parameter

Choice

Description

VER

Value

Displays the current software version.

BAT

Value

Displays the current battery level.

VALUE

Value

Displays the actual value.

A2D

Value

Displays the actual raw counts of the scale.

Table 7-1. Test Menu

DEF

CALIB

PROG

DONE

TEST

VER

BAT

VALUE

A2D

1.83

Value

Value

Value