Installation Manual

Compression Load Cell PR 6201

Sartorius

EN-21

6

Preparing for Calibration

Note:

For calibration of the measuring system, please refer to the manual of the weighing
instrument.

Caution!
A warming-up time of at least 30 minutes for the load cell is required before calibration

of the weighing installation.

6.1

Mechanical Height Adaptation

Before starting calibration it has to be ensured that the load cells are equally loaded. Therefore a height
adaptation has to be performed.
• Place the dead load (e.g. empty hopper) on the load cells.
• Energize the load cells by means of a stabilized voltage of e.g. 12 V DC.
• Measure the output voltage of each individual load cell by means of a high-accuracy digital voltmeter and

compare the individual values.

The load cell output with the lowest voltage must be increased by putting shims between mounting plate and
weighing construction.
• Lift weighing facility near the load cell.
• Put a thin, deburred sheet of metal (0.5 mm to 2 mm thick) between the upper mounting plate and

weighing construction.

• Measure the output voltage of the load cells again and correct the height of this load cell or another

one.

7

Fault finding hints

The following hints will enable a technician for a first diagnostic or help in case of wrong or not
reproducible weighing results after commissioning and calibration.

7.1

Visual check

Weighing facility

Are all pipes, hoses, and cables connected as flexibly as possible?

Are devices which are rigidly connected to the weighing facility in direct
contact with the surrounding installation?

Is there friction between the weighing facility and its surrounding (e.g. dirty
apertures etc.)?

Junction box

Has moisture penetrated the junction box?

Do the screw connections show a good contact?

Connecting cable

Is the sheath damaged?

Has moisture penetrated the cable end?

Mounting kit

Is there a contact between lift-off protection and the weighing facility
itself?

Are the constrainers stuck?

Load cell

Does the load cell stand vertically?

Is the housing damaged?

Is the sheath of the cable damaged?

Has moisture penetrated the cable end?