2 checks with multimeter, 1 check zero output signal of load cell, 2 check the strain gauge bridge – Rice Lake PR6201 Zinc-Plated Steel Star Mount Kits User Manual

Page 24: 3 check insulation impedance of load cell, 4 check insulation impedance of connecting cable, Compression load cell pr 6201 installation manual

Advertising
background image

Compression Load Cell PR 6201

Installation Manual

EN-22

Sartorius

7.2

Checks with Multimeter

7.2.1

Check Zero Output Signal of Load Cell

• Unload load cell.
• Disconnect the load cell outputs.
• Check whether the output voltage without load is within the limits.

Type

Zero output signal

L

0 mV ±0,02 mV/V

D1/N/C3

0 mV ±0,01 mV/V

(at PR 6201/54..

0 mV ±0,02 mV/V)

LA

3,2…4 mA
GAIN connected to +Supply Voltage 24 ±4 V, see chapter 5.2.2.1

7.2.2

Check the Strain Gauge Bridge

Note:

Not for load cells type LA.

• Do not exceed the test voltage.
• Check whether the values of the resistors are within the permissible limits.
Maximum test voltage

Standard industrial

32 V DC

Intrinsically safe version (PR …/..E)

25 V DC

Type

Input impedance
(red core, blue core)

Output impedance
(green core, grey core)

L

650

Ω +50 Ω

610

Ω ±3 Ω

D1/N

650

Ω ±6 Ω

610

Ω ±1 Ω

C3

650

Ω ±6 Ω

610

Ω ±0,5 Ω

7.2.3

Check Insulation Impedance of Load Cell

Note:

Not for load cells type LA.

• Never apply the test voltage between the cores of the load cell cable (danger of destroying the load

cell).

• Insulate the load cell cores.

maximum test voltage

standard industrial

100 V DC

Intrinsically safe version (PR …/..E)

500 V AC

insulation impedance

core - housing

>5000 M

core - screen

>5000 M

screen - housing

>5000 M

7.2.4

Check Insulation Impedance of Connecting Cable

• Disconnect connecting cable and load cells.
• Insulate the cores of the connecting cable.

insulation impedance

core - core

≥600 MΩ × km

core - screen

≥600 MΩ × km