Rice Lake RD-300 Remote Display User Manual

Page 13

Advertising
background image

Installation

11

Enter the baud rate from the table listed below.

Select the interface type, RS-232 or 20 mA with SW2-8.

Note:

IF YOU ARE USING TOLEDO, MASSTRON, FLEX-WEIGH, OR ANALOGIC (or IQ plus 310 IN ANALOGIC MODE) METER
SETTINGS, USE THE FOLLOWING SWITCH DATA

Toledo, Flex-Weigh, Analogic, and Masstron meters all transmit data continuously. They also have special
information available in their formatting which allow the RD-300/RD-232 to perform additonal useful
functions. With these meters, the timeout function is always enabled. SW2-4,5,6 and 7 have the following
meanings:

The RD-300/RD-232 will display the data selected by SW2-6 and 7 regardless what is showing on the scale
indicator.

IF YOU ARE USING ANY OTHER METER SETTING, USE THE FOLLOWING SWITCH DATA ON PAGE 12.

Baud Rate

SW2-1

SW2-2

SW2-3

38,400

On

On

On

19,200

Off

On

On

9,600

On

Off

On

4800

Off

Off

On

2400

On

On

Off

1200

Off

On

Off

600

On

Off

Off

300

Off

Off

Off

Table 2-2. Baud Rate Settings

Input Type

SW2-8

20 mA CL

On

RS-232

Off

Table 2-3. Interface Type

Retransmit Data Format

SW2-4

Condec data format

On

As received

Off

Table 2-4. SW2-4 Data Format Settings

Display on Time Out

SW2-5

“No sig”

On

Last data received

Off

Table 2-5. SW2-5 Time Out Settings

Displayed Weight Mode

SW2-6

SW2-7

Current displayed weight

On

On

Gross weight only

Off

On

Net weight only

On

Off

Tare weight only

Off

Off

Table 2-6. SW2-6/7 Displayed Weight Settings