Other meters may transmit data intermittently, slowly, or in a demand mode where the data is only sent once.
Enabling the time out function causes the RD-300/RD-232 to do the following. If new data hasn't been received
for approximately two seconds, then the display shows what is selected by SW2-5, either the last valid weight or
all dashes. If no valid data is received after ten seconds, the RD-300/RD-232 initiates a power on reset, clear all
data, read the switches, and waits for new data. For these meters SW2-4,5,6 and 7 have the following meanings:
Time Out Enable
SW2-7
Time out counter disabled (cn dis)
Off
Time out counter enabled (cn ena)
On
Table 2-7. SW2-7 Time Out Enabled Setting
Display on Time Out
SW2-5
“No sig”
On
Last data received
Off
Table 2-8. SW2-5 Time Out Settings
Universal KG Character Type
SW2-6
A single character is sent, K, k
Off
KG, Kg, or kg is sent
On
Table 2-9. SW2-6 Character Type Settings
Retransmit Data Format
SW2-4
Condec data format (rd con)
On
As received (rd rec)
Off
Table 2-10. SW2-4 Data Format Settings