RoughDeck BDP Installation/Operation Manual

11

6.0

Service Information

The following sections describe basic service and maintenance procedures.

6.1

Troubleshooting Guide

Table 6-1 lists some of the common problems and their suggested solutions.

6.2

Periodic Maintenance

The space between the platform side and the portable deck assembly, and the surface beneath the platform must

be periodically cleaned to prevent debris build up. More frequent cleaning of these areas is necessary with scales

permanently mounted in pits.
Do not attempt to use scales with load cells that are not hermetically-sealed in washdown applications. Water

damage is a common cause of failure in non-hermetically-sealed load cells. Use care with high pressure steam

washdowns for hermetically-sealed load cells. The steam may not damage the load cells, but the elevated

temperatures may cause incorrect readings until the unit cools to room temperature.

Problem

Symptom Description

Solution

System does not
operate - no display

Power disconnected

Check and reconnect

Indicator fuse blown

Replace the fuse. Check for cause

Interface cable cut or disconnected

Repair

Signal leads incorrectly installed at the
indicator

Install according to indicator installation
manual

Display stays at zero

Load cell connections faulty

Check cable connections in junction box and
at indicator

Indicator faulty

Service indicator

Erratic weight
readings

Vibration near scale

Remove source of vibration or move scale

Platform not level to within 1/4 inch

Level scale by adjusting feet or shimming if
necessary

Load cell or cable water damage

Replace

Debris under load cells or platform

Clean

Indicator faulty

Use simulator to test indicator for stability.
Service indicator

Consistently high or
low weights

Indicator not properly adjusted to zero

Zero the indicator according to the indicator
manual

Platform binding

Obtain adequate clearance for free platform
movement

Indicator not calibrated

Calibrate according to indicator manual and
See Section 5.3 on page 10

Load cells faulty

Test and replace load cells if necessary

Feet touching deck underside

Adjust feet downward to provide clearance

Table 6-1. Troubleshooting Table