RoughDeck BDP Installation/Operation Manual

5

3.2

Access Ramps

Optional access ramps for the RoughDeck BDP barrel scale come in both mild steel and stainless steel versions

and have a standard treaded plate. They are designed to bolt to the floor, with built-in mounting plates that attach

to the scale feet. When used with access ramps, side movement of the scale is automatically eliminated, and no

other mounting plates are necessary.

Access ramps can only be attached to the scale on one of the two scale sides that are perpendicular to the

longitudinal axis of the load cells and that are perpendicular to the side rails. For example, the scale shown below

could have an access ramp on the left side as shown, and/or on the right side. The top and bottom sides, however,

will not accept the ramp mounting plates.

Figure 3-2. Optional Access Ramp

3.3

Electrical Interface to Indicator

Ten feet of 6-wire cable to connect the scale to the weight indicator is supplied with each scale. The junction box

is easily accessible through a top access plate located on the top of the RoughDeck BDP. Use the following steps

to wire up the junction box:

1. Remove the two screws using an 1/8 inch allen wrench to remove the scale junction box cover plate.
2. Slide the junction box assembly out of the deck.
3. Open and remove the top of junction box.
4. Push the cable end into the junction box through a cord grip.
5. Connect the wires to the indicator terminal (Figure 3-3) as shown in Table 3-1.
6. Pull out excess cable and tighten the cord grip to hold the cable snugly.

Cable Color Code

Junction Box

Red

+ Excitation

Black

- Excitation

Green

+ Signal

White

- Signal

Bare

Shield

Yellow

+ Sense

Blue

- Sense

Table 3-1. Junction Box Connections