12 SCT-30 Weight Transmitter Manual

Tot. bytes = 3+2*No. registers+2

where: No. of registers = number of Modbus registers to be read, starting

from register 1 address;

No. of bytes = number of following data bytes;

In the event of a string received correctly but not executable, the slave responds with
an EXCEPTIONAL RESPONSE. The FUNCTION field is transmitted with the msb

at 1.

The communication strings are controlled by CRC (Cyclical Redundancy Check).

In case of a communication error the slave will not respond with any string. The
master must allow for a time-out before response reception and if no response is
received, it infers that a communication error has occurred.

Registers and Values

The instrument ModBus registers may only be read.

H - L: high half and low half - respectively, making up the DOUBLE WORD value.

RESPONSE

Address

Function No. bytes Register1 Register 2

2 bytes

A

0x03

0x04

0x0064

0x00C8

CRC

EXCEPTIONAL RESPONSE

Address

Function

Code

2 bytes

A

Funct + 0 x 80

CRC

CODE

DESCRIPTION

1

ILLEGAL FUNCTION (Function not valid or not supported)

2

ILLEGAL DATA ADDRESS (The specified data address is not available)

3

ILLEGAL DATA VALUE (The data received have no valid value)

REGISTER

DESCRIPTION

40007

Status Register

40008

H Divisions

40009

L Divisions