Calibration

5

3.With the

ZERO

key held, press and release the

TARE

key, then press and

release the

POWER

key twice.

The SPEC number is displayed.

4.To increment the SPEC setting value, press the

POWER

key.

Refer to Table 3-1.

5.Press the

ZERO

key to save data and increase the SPEC count.

6.Press the

ZERO

key to advance to the next SPEC.

7.Press the

TARE

key to store SPECs and return to weigh mode.

SPEC

Number

Bit 2

Bit 1

Bit 0

00

Decimal point position
000 - None
001 - 2nd digit (0.0)
010 - 3rd digit (0.00)
011 - 4th digit (0.000)
100 - 5th digit (0.0000)
101 ~ 111 - Not used

01

Minimum display
00 - 1
01 - 2
10 - 5
11 - 10

Net/gross multi-interval
0 - Gross
1 - Net

02

Resolution Selection
00 - 1/2000
01 - 1/3000
10 - 1/6000
11 - 1/7500

Weight single interval or
multi-interval
0 - Single interval

03

Decimal point figure for
numeral
0 - Period
1 - Comma

Negative weight display
mask
0 - Minus gross > 9e
1 - Minus gross weight

IR (inter resolution) mode
protected by SPAN SW
(SPAN switch must be
turned on for calibration)
0 - No
1 - Yes

04

Start range (amount of weight that can be on the
scale when it is turned on; if there is more weight
than allowed, the display will show all 8s)
00 - ±10% F.S. 10 - ±3%F.S.
01 - ±5% F.S. 11 - ±2%F.S.

Re-zero range
0 - ±2% F.S.
1 - ±3% F.S.

05

Not NTEP

Approved

Manual tare cancellation
0 - Allow
1 - Prohibit

Tare subtraction
0 - Allow
1 - Prohibit

Tare accumulation
0 - Allow
1 - Prohibit

Table 3-1. SPEC Number And Bit Explanation (Defaults Are Bold)