3

3.0

Calibration

To enter calibration mode, remove the screw on the bottom of the scale as shown

in Figure 3-1. Use a small tool to press the span setup switch located under the

screw. Setup will time out to verify the switch is set to the

ON

position. Then,

perform a span switch state check (see Section 3.1).

Remove Screw

Figure 3-1. Setup Switch

3.1

Software Version and Span State Check

1.With the unit in weigh mode, press and hold the

ZERO

key.

2.With the

ZERO

key held, press and release the

POWER

key three times.

3.The display will show the software version for two seconds, then show the

current span switch state (

S-ON

or

S-OFF

).

After three seconds, the unit will return to weigh mode.

3.2

Performing a Calibration

The Versa-portion RLP-S/SM must be calibrated in grams or kilograms, and

SPEC 01 and 02 must be set for the gram value (or kg in the SM version). See

Section 3.4.

1.With the unit in weigh mode, press and hold the

ZERO

key.

2.With the

ZERO

key held, press and release the

POWER

key twice, then press

and release the

TARE

key.

The display will show CAL00.