Calibration report – Torbal AGCN200 User Manual

Page 17

Advertising
background image

Instruction Manual - TORBAL AGCN Series

3. Press the T Key again when “CAL °C” is displayed.

4. Temperature interval settings will begin to display sequentially. The current temperature interval setting will

be marked with “o” located in the lower left corner of the display.

5. Automatic calibration can be set to 0.5°C, 1.0°C, 1.5°C, and 2.0°C (i.e., if set to 2.0°C, calibration will be

performed every time the temperature changes by 2.0°C).

6. To select a temperature interval, press the T Key when the desired setting is displayed.

To disable temperature calibration, press the T Key when OFF is displayed.

8.1.5.Calibration Report

After calibration set-up and adjustment have been completed, a calibration report can be printed. The

calibration report includes calibration data such as the current as well as factory internal and external weights. To
print the calibration report, follow these steps:

1. Press MENU to enter the main menu.

2. When command “CALIBR” is displayed, press the T key.

17