Torbal AGCN200 User Manual

Page 9

Advertising
background image

Instruction Manual - TORBAL AGCN Series

3. Once the pan has been installed, level the scale by adjusting the rear feet until the level indicator shows the

“air bubble” is in the center position of the sight glass. The level indicator is located in the rear left side of the
scale.

4. After leveling the scale, you may plug the AC adaptor to the AC adaptor socket located in the rear of the scale.

5. When the AC adapter is plugged into the wall outlet, the scale will turn “ON” automatically, initialize itself,

perform automatic internal calibration and enter weighing mode.

6. To put the scale into standby mode, leave the AC adaptor plugged into both the scale and the wall outlet and

press the Power “OFF” button (

). The “OFF” indicator will light up in the upper left corner of the

display signaling the scale is in standby mode.

9

Advertising
This manual is related to the following products: