4 interface and usb port configuration – Torbal FA Series User Manual

Page 23

Advertising
background image

- 23 -

Instruction Manual

– FA Series Force Gauge

12.4 Interface and USB Port Configuration


The Interface menu is designed to allow you to change the format and rates at which information is transmitted
to a computer. You must connect the force gauge to the computer using the USB port. To access this function
go to CONFIGURATION in the Main Menu.

* To connect the force gauge up to a computer it is recommended that TORBAL FM is installed on the
computer. With this program the user can view real-time results and measurements from the force
gauge. TORBAL FM also allows the user to fully control the force gauge from the computer. The
TORBAL FM settings must match those of the force gauge.


The optimal settings from transmitting information to a
computer are as follows:

-Baudrate (transmission and receiving rate): 4,800, 9,600,
19,200, 38,400, 57,600, 115,200 bps.

-Bits
(the number of bits which define a character): 7 or 8
bits.

-Parity (refers to the alteration of the process that the bits are sent and received in): none, odd, even.

-Sending (the transmission method during measurement):
Normal

Waits for stable results after pressing PRINT before sending.

NO STB

Sends results regardless of stable results after pressing PRINT.

AUTO STB

Automatically sends after the result has stabilized.

CONTIN.

Sends continuous transmissions (approx. every 0.1s).



Description of the data transmission protocol when working with a computer (LonG):

The scales transmit the result as follows (8 bits, 1 stop, no parity, 4,800 bps):
Computer

Gauge: initiating signal S I CR LF (53 h 49 h 0Dh 0 Ah),

Gauge

Computer: gauge indication according to the following format (16 bytes)


Description of individual bytes:

byte

1

- “-“ or space

byte

2

- space

byte

3-4

- digit or space

byte

5-9

- digit, comma or space

byte

10

- digit

byte

11

- space

byte

12

- k, l, c, p or space

byte

13

- g, b, t, c or %

byte

14

- space

byte

15

- CR

byte

16

- LF

INTERFACE

1. Baudrate 57600
2. Bits 8-bit
3. Parity none
4. Sending NORMAL

5. Exit

Advertising