Chapter 11 Specifications

Welch Allyn Acuity Central Monitoring System

Clicking Arr/ST Relearn or RELEARN

Clicking Analyze Pacers

Clicking Single ECG

Note: A RELEARN button instead of
an Arr/ST Relearn button indicates
that the system does not include ST
Analysis or the ST Analysis Setup
window.

System relearns the patient’s normal
ECG rhythm and uses the relearned
rhythm as a guideline to identify
arrhythmia events (arrhythmia relearn)
and to analyze deviation of the ECG ST-
segment offset (ST relearn).

In Virtual Monitor and Waveform
Window: A visual RELEARN alert
occurs, HR is taken from the patient
monitor instead of from the arrhythmia
software, and real-time waveforms
display as usual.

In ST Analysis Setup window: The
waveform slice displays NO VALID
BEATS, and the Acuity System can take
up to 30 seconds to update the
waveform.

Relearn and Learning Done event
markers (

▼) appear in most review

windows and some printouts.

Table 35. Setup window

Arrhythmia Alarms

Features

Customizable default

Factory default