Directions for use

Chapter 4 Adjust monitoring settings and patient information

73

Set analysis preferences for a patient’s ST analysis

To set analysis preferences for ST analysis

1.

In the patient’s Virtual Monitor, click Setup, ST Analysis.

The ST Analysis Setup window appears.

2.

Leads: In the Lead column, check leads to be used for ST Analysis.

Depending on the patient monitor, some of these choices cannot be deselected.

3.

Measurement Offset: Click to set the offset from the J-point. This is the point at
which the ST measurement is made.

If you click Auto, the Acuity System selects the offset based on the patient’s heart
rate.

J-points (green)