Brookfield DV-III Ultra Rheometer User Manual

Page 116

Advertising
background image

Brookfield Engineering Laboratories, Inc.

Page 116

Manual No. M/98-211-B0104

Brookfield Engineering Laboratories, Inc.

Page 116

Manual No. M98-211-E0912

Calibration Procedure for LV (#1-3) and RV,HA,HB (#2-6) Brookfield spindles

:

Note:

The LV #4 (#64) and RV, HA, HB #7 spindles have been omitted from this

procedure. Brookfield does not recommend the use of these spindles to perform

a calibration check on your instrument. Reasons pertain to the small amount

of surface area that makes contact with the viscosity standard, the difficulty of

establishing the immersion mark precisely and the need for precise temperature

control at 25°C in the immediate vicinity of the spindle.

Follow these steps using one of the recommended spindles to verify calibration of your

instrument.

1) Place the viscosity standard fluid (in the proper container) into the water bath.

2) Lower the DV-III Ultra into measurement position (with guard leg if LV or RV series Rheometer

is used).

3) Attach the spindle to the Rheometer. If you are using a disk shaped spindle, avoid trapping air

bubbles beneath the disk by first immersing the spindle at an angle, and then connecting it to

the Rheometer.

4) The viscosity standard fluid, together with the spindle, should be immersed in the bath for a

minimum of 1 hour, stirring the fluid periodically, prior to taking measurements.

5) After 1 hour, check the temperature of the viscosity standard fluid with an accurate

thermometer.

6) If the fluid is at test temperature (

+

/- 0.1

°

C of the specified temperature, normally 25

°

C),

measure the viscosity and record the Rheometer reading.

Note:

The spindle must rotate at least five (5) times before readings are taken.

7) The viscosity reading should equal the cP value on the viscosity fluid standard to within the

combined accuracies of the Rheometer and the standard (as discussed in the section entitled,

Interpretation of Calibration Test Results).

Calibration Procedure for a Small Sample Adapter

When a Small Sample Adapter is used, the water jacket is connected to the water bath and the

water is stabilized at the proper temperature:

1) Put the proper amount of viscosity standard fluid into the sample chamber. The amount varies

with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction

manual.)

2) Place the sample chamber into the water jacket.

3) Put the spindle into the test fluid and attach the extension link, coupling nut and free hanging

spindle (or directly attach the solid shaft spindle) to the DV-III Ultra.

4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test

temperature.