IV.8.3

Data Page

Figure IV-60

Use this page to load, save, print, and export data taken with the DV-III Ultra Rheometer. The file

name (including path) of the data currently loaded is displayed at the top of the page.

The test parameters used to collect the data are displayed on the left of the page in the box labeled

“Parameters Used To Collect Data”.

Data is displayed on the right of the page in the box labeled “Data/Results”. All recorded data

points are displayed in the table.

Time

Elapsed time since the beginning of the run cycle is shown here in units of

hours, minutes, seconds, and thousandths of seconds (

H:MM:SS.SSS).

Data is taken at an interval equal to the Base Increment parameter.

% Torque

This field displays actual measured % Torque value

Delta % Torque

This field displays the difference between the current % Torque value and

the previous % Torque value. As the sample approaches its yield point,

the Delta % Torque should begin to decrease. If the Torque reduction for

the test is 100%, the yield point occurs when the % decrease in torque =

100.

Stress

This field displays the calculated stress for each data point. The final Stress

value (i.e. the Stress at the yield point) is the resultant yield stress. See

Appendix D - Calculations - for the equation used to calculate stress.