6 Operation

45

K-439 Operation Manual, Version A

➡ The samples are now digested. Follow-up actions are described in section 6.10.

!

DANGER

Risk of death or serious injuries by toxic and explosive peroxides.

Use only original glassware and funnels

Do not use damaged filter funnels

Only add peroxides directly into the hot sample via filter funnel

Exchange worn out or defective parts before digestion

Handle peroxides in a fume hood

Do not bring peroxides into contact with highly combustible material

Dispenser
with peroxide


Filter funnel

Rubber
GUKO

The funnels are placed decentralized (shifted out of the middle) above the sample tubes to prevent
peroxide dripping directly into the hot sample. The filter funnel ensures that the peroxide can not flush
into the hot sample to avoid dangerous deflagration of the peroxide and the sample.