8 Troubleshooting

57

K-439 Operation Manual, Version A

Malfunctions and their remedy

Malfunction

Possible cause

Remedy

System does not heat

No voltage

Insert mains plug, examine plug for
damage, check power switch

Mains switch is turned off

Switch must light up green, when
system is powered up

Fuse has been activated

Exchange fuse(s), see section 8.3

Heating elements are defective

Contact Buchi customer service to
replace defective component(s)

Heating elements are overheated–
overtemperature safety switch is
activated

Contact Buchi customer service for
repair of defective component(s)

Fumes are escaping from the
suction module or its assemblies
into the fume hood

Scrubber or water jet pump

not switched on or functioning
correctly

Defective or clogged sealing,

glassware or tubes

Stop the digestion process

Check and clean all parts

See manuals of the Scrubber / water

jet pump on how to increase
suction power


Malfunctions or errors which are not listed in the table are to be corrected by a Buchi trained technical
engineer who has access to the official service manuals. In this case, please refer to your local Buchi
customer service agent.

8 .2

Diagnostics

The K-439 software enables the user to carry out a service test (OQ) and to view actors, sensors,
operating hours and unit information.
To open the Diagnostics menu select Main Menu > Diagnostics. The following screen appears:

8 .2 .1

Actors

This submenu enables the user to view the heating output of the heaters and the status of the
Scrubber B-414 and the backlight:

Heaters: xx %

xx °C

Scrubber: On/Off

Backlight: On/Off