8 - Maintaining the instrument

179

 Set oxygen dosing parameters (method) for conditioning and calibration samples according to

the following list:

Weight 2 - 20 mg:

O

2

dosing time 110 sec.

O

2

flow = 50 ml/min.

O

2

cut-off threshold = 30 %.

Weight 20 - 100 mg:

O

2

dosing time 110 sec.

O

2

flow = 100 ml/min.

O

2

cut-off threshold = 30 %.

Weight 100 - 200 mg:

O

2

dosing time 120 sec.

O

2

flow = 150 ml/min.

O

2

cut-off threshold = 30 %.

Weight 200 - 300 mg:

O

2

dosing time 120 sec.

O

2

flow = 200 ml/min.

O

2

cut-off threshold = 30 %.

Weight 300 - 400 mg:

O

2

dosing time 120 sec.

O

2

flow = 250 ml/min.

O

2

cut-off threshold = 30 %.

Weight 400 - 500 mg:

O

2

dosing time 120 sec.

O

2

flow = 300 ml/min.

O

2

cut-off threshold = 30 %.

 Start the analysis process by selecting System > Auto run.

 Check whether the field "MFC O2" on the screen displays the expected setpoint value during the

combustion time. See Values for oxygen dosing (on page 137).

 Observe the measuring results. If the N blank value is not below 50 after stabilization: Stop

continuous analysis via System > Single. Find out the reason of the increased blank value and
correct it. Cf. Types of blank value determination and their settings (on page 143).

 Save the original values. See Data backup and printing (on page 158).

Phase 2: Ascertaining appropriate calculation methods

Proceed as follows:

 Execute the following command: Math. > Calibration. If sequential blank value determination is

activated a message window opens showing the corresponding information.

 Click Yes to continue. The "Configure calibration" dialog appears.

 Define the following settings:

Check the box next to "Two calibration ranges".

Enter 1 as the polynomial degree for the lower and upper range. This corresponds to linear
calculation.