5 Putting into operation

21

E-812/816 Hot Extraction Operation Manual, Version B

5

Putting into operation

This chapter describes how the instrument is installed and gives instructions on initial startup.

NOTE
Inspect the instrument for damages during unpacking. If necessary, prepare a status report immedia-
tely to inform the postal company, railway company or transportation company.

Keep the original packaging for future transportation.

5 .1

Installation site

Put the instrument on a stable, horizontal surface. Consider the maximum product dimensions and
weight. Conform to the environmental conditions as described in section 3.2, technical data.

Installation prerequisites:
• Do not place any objects on top or below the instrument.
• The instrument must be installed with 5 cm clearance from any other objects or walls to allow

sufficient cooling.

• Do not store containers, chemicals or other items behind the instrument.

!

WARNING

Death or serious injuries by use in explosive environments.
• Do not operate the instrument in explosive environments
• Do not operate the instrument with explosive gas mixtures
• Before operation, check all gas connections for correct installation
• Directly withdraw released gases and gaseous substances by sufficient ventilation

!

CAUTION

Risk of minor or moderate injury by heavy weight of the instrument.
• Consult three further persons to transport the instrument
• Do not drop the instrument
• Place the instrument on a stable, even and vibration-free surface
• Keep limbs out of crushing zone

NOTICE

Risk of instrument damage by liquids or mechanical shocks.
• Do not spill liquids over the instrument or its components
• Do not move the instrument when it is loaded with sample liquid
• Do not drop the instrument or its components
• Keep external vibrations away from the instrument
• Safely attach the instrument to the bench in earthquake prone regions
• Do not operate the instrument without the protection cover installed at the front