8 troubleshooting, 1 malfunctions and their remedy, Malfunctions and their remedy – BUCHI Extraction Unit E-816 HE User Manual

Page 47: 8 .1 malfunctions and their remedy

Advertising
background image

8 Troubleshooting

47

E-812/816 Hot Extraction Operation Manual, Version B

8 Troubleshooting

This chapter helps to resume operation after a minor problem has occurred with the instrument. It lists
possible occurrences, their probable cause and suggests how to remedy the problem.
The troubleshooting table below lists possible malfunctions and errors of the instrument. The operator
is enabled to correct some of those problems or errors by him/herself. For this, appropriate corrective
measures are listed in the column “Corrective measure”.
The elimination of more complicated malfunctions or errors is usually performed by a BUCHI technical
engineer who has access to the official service manuals. In this case, please refer to your local BUCHI
customer service agent.

8 .1

Malfunctions and their remedy

Table 8-1: General malfunctions and their remedy

Malfunction

Possible cause

Corrective measure

Instrument does not work

Incorrect or no mains connection

Connect the instrument correctly to the mains
supply

Instrument not switched on

Switch on the instrument

Method cannot be started

Service mode active

Change to Extraction mode

Rack is in upper position

Lower the rack

Solvent tank is full

Empty the tank

Protective shield is open

Close the protective shield

Software does not show the main
screen

Go back to the main screen

Method stops

Cooling water flow is too low

Check the cooling water connection and all
hoses for bends and remove them, if neces-
sary

Cooling water sensor is defec-
tive (may be the feeding water is
contaminated with algae, lime, etc.)

Contact the BUCHI customer service

Evaporation of extraction
solvent

Seals wrong, deformed or damaged

Exchange the seals

Spring deflection downwards

Contact the BUCHI customer service

Incorrect positioning of the beaker/
glassware and seals

Contact the BUCHI customer service

Boiling retardation

Incorrect positioning of the beaker

Place the beaker directly onto the heating
plate

No anti-bumping granules were used Put anti-bumping granules into the beakers
Too long waiting time between
preparation of the instrument and
start

Start the instrument quickly after solvent was
added