either of the two following calibration intervals.
Performance Based Method
Determine need to calibrate by performing weekly check-runs of certified

viscosity reference standards. It is recommended the results of these runs

be recorded and compared against the certified viscosity value printed on

the standard bottle. When these check-runs consistently exceed the required

precision (typically > 2 or 3%), then the instrument should be calibrated. This

method requires more user interaction, but affords the least amount of risk

that the instrument could be used when out-of-calibration.
Periodic Interval Method
On a 12-month basis, the instrument is returned to CANNON

®

Instrument

Company for calibration and certification regardless of the prior usage and

performance. This method requires the least amount of user interaction with

the instrument, but does afford some risk as errant calibrations (if present) will

only be corrected once every 12 months.

Verifying Temperature

Periodically, the temperature should be verified with

a digital reference thermometer. The temperature

should be recalibrated if the displayed temperature

does not agree with the reference thermometer.
Note: Please contact CANNON

®

Instrument

Company for details on suitable reference

thermometers.
The user may access Temperature

Calibration Screen from the Ready to Test

Screen by tapping the temperature field 3 times within 2 seconds. The

Temperature Calibration Screen will appear in which the user may

enter the temperature as read from the reference thermometer to the nearest

0.01°C. This may need to be repeated if the temperature still does not agree

Maintenance & Calibration

Temperature Calibration

Screen