Unpack your SimpleVIS

®

Viscometer

Remove all packing materials from the components. The SimpleVIS

®

Viscometer should be placed on a stable laboratory table or bench.
1. Verify that you have received all components for the SimpleVIS

®

Viscometer. Report missing items to CANNON

®

Instrument Company

immediately.

2. Inspect each component for signs of damage. Report damages to the

shipper and CANNON

®

Instrument Company immediately.

3. Retain all packing materials until the instrument is connected and

functioning properly. If any component(s) must be returned to CANNON

®

Instrument Company, the damaged item(s) should be packaged in the

included shipping case.

Setting up your SimpleVIS

®

Viscometer

Complete the following rear panel electrical connections:
1. Plug the AC/DC power adapter into the power connector on the rear

panel of the instrument. The connector will only go in one way correctly.

Make sure the power switch on the AC/DC adapter is off.

2. Plug the power cord into the AC/DC adapter. Plug the other end of the

power cord into an appropriate AC Mains outlet matching the voltage

requirements indicated on the AC/DC adapter.

3.

Do not turn on the AC/DC adapter switch at this time.

(Optional printer: Plug the 9 pin serial printer cable into the RS232-C port on

the rear panel of the instrument. Then plug the other end into the printer.)
Complete the following tubing connections:
1. Locate the glass waste bottle and securely (but not overly) tighten the

bottle cap to ensure an air-tight seal. Place the bottle in the holder on the

top rear of the SimpleVIS

®

Viscometer.

2. Connect both pieces of the1/8” transparent tubing into the bottle cap.
3. Make sure that the waste line extends into the cap further than the