Running a test – Cannon Instrument SimpleVIS User Manual

Page 7

Advertising
background image

Print the results (only if

optional printer is installed)

1. Press

the

Print button on the

display screen to printout the test

results to an attached serial printer

(optional accessory).

2. You may skip the print option and

go directly to the wash cycle. See

the next section.

Wash Cycle

1. Place the tip of the solvent bottle

along the opening of the glass

capillary in the sample/solvent port and gently squeeze until the cup is

filled with solvent to clean any oil residue that might be left over from

injecting the previous sample.

Be careful to consider bath temperature (s) and potential solvent

hazards when selecting solvents for use with the SimpleVIS

®

Viscometer.

Observe appropriate safety precautions when handling solvents.
Note: Heptane (technical grade) is compatible with the

SimpleVIS

®

Viscometer.

2. Press

the

Wash button on the display screen to start the wash cycle.

3. Wait for the solvent in the cup to be drawn down the tube.
4. Place the tip of the solvent bottle directly into the glass capillary for

approximately one second (no need to squeeze the bottle, the pump

will pull the solvent in) and then remove for several seconds to allow the

solvent to be drawn through the viscometer tube to the waste bottle. The

display screen will prompt the user when to inject and remove the solvent

bottle. Repeat this step until the display screen alerts the user to press the

Dry button to start the dry run cycle.

Running a Test

Result Screen

7