The ~MAINS symbol indicates instructions or connections for the AC power

supply. The AC Power input must match the electrical specifications listed on the

label on the rear panel of the instrument. The supplied AC Mains power cord

must be attached to the connector labeled ~MAINS. This connection serves as a

means of disconnect and should be readily accessible.
The (O) symbol indicates the OFF position for the electrical switches for your unit

(AC Mains or accessories).
Routine SimpleVIS

®

Viscometer operation may require the use and

handling of hazardous chemicals and solutions. CANNON

®

Instrument Company

strongly urges the operators and technicians working with the SimpleVIS

®

Viscometer to take proper safety precautions when working with these materials.

These safety procedures can be found in the Material Safety Data Sheets which

accompany the solutions.

SimpleVIS

®

Viscometer Specifications/Compliance

Product Info.

Model: SimpleVIS

®

Viscometer (Cat. No. 9725-S10)

Applications

ISO VG Lubricants, SAE Engine and Gear Oils

Methodology

Kinematic Viscosity, ASTM D7279 or ASTM D445-Modifi ed

Test Temperatures

40 or 100 °C - user selectable

Viscosity Range

40°C: 10 – 700 mm

2

/s (cSt)

100°C: 5.5 – 200 mm

2

/s (cSt)

Accuracy

< 3% of measured value, calibrated

Repeatability

< 2% of measured value, typical

Sample Material Type

Mineral Oils, Synthetic Oils, Formulated Motor/Gear Oils

Solvent Compatibility

Internal wetted components compatible with n-heptane only

Volume Requirements

Sample: 450 – 500 μL Solvent: 10 – 15 mL per test

Dimensions

128 mm wide x 356 mm deep x 304 mm high (5x14x12”)

Weight

SimpleVIS

®

Viscometer unit: 4.6 kg (10.2 lbs)

Shipping Weight

12.3 kg (27 lbs) with all accessories

Input Power

SimpleVIS

®

Viscometer Operation Voltage: 11 – 13 VDC @ 5A peak

AC/DC Adapter

Power Requirements

100~230 Volts, 50/60 Hz, 100 Watts

Operating Conditions

15°-30°C, 10%-90% R.H. non-condensing, Installation Category II, Pollution degree 2

Compliance

CE Mark: EMC Directive (2004/108/EC); Low Voltage Directive (2006/95/EC);

HI-POT (1900 VDC, 60 sec.); ROHS