2139 High Tech Road • State College, PA • 16803 • USA

switch ON to maintain temperature control. When the PH switch has been
turned on, the PH - HEAT LED will light.

When the bath temperature is approximately that selected with the T -
SELECT
switch, the LED above the TA controls will start to blink as
the heater duty cycle is automatically adjusted to stabilize temperature.

3. Release the lock on the TA - VARIABLE dial.

4. Read the thermometer in the CT-500 bath to determine the actual

temperature of the bath. Make sure that both the thermometer and the
bath temperature have stabilized before taking a reading.

5. Turn the TA - VARIABLE dial to adjust the temperature. Turn the

dial clockwise to increase the bath temperature. Turn the dial coun-
terclockwise
to decrease the bath temperature. After each adjust-
ment, allow several minutes for the bath temperature to stabilize.

6. When the desired temperature has been reached, re-lock the TA -

VARIABLE dial.

Limit Control

After the bath has reached the desired temperature, set the HIGH
TEMPERATURE LIMIT CONTROL (LC)
. This will prevent the bath
from overheating if a malfunction occurs.

Adjusting the High Temperature Limit Control

NOTE

During this procedure, the temperature of the bath may change slightly,
but will quickly recover.

1. Using a screwdriver, slowly turn the LC - TEMPERATURE

ADJUST control counterclockwise until the OVER TEMP message
lights up on the LC - PUSH TO RESET button.

2. Turn the control clockwise approximately ¼ to ½ turn.

3. Push the LC - PUSH TO RESET button. If the bath doesn't recover,

repeat step 2 and try again.

NOTE

If the desired temperature is within 2 or 3°C of a temperature listed on
the T - SELECT dial, it may be possible to attain the temperature using
the TA - FIXED dial. Because this dial has a finer adjustment than the
TA -VARIABLE dial, the final temperature can be more easily obtained.
To try this alternative method, follow the previous instructions, using the
TA - FIXED dial instead of the TA - VARIABLE dial.

Advertising