Viscosity standards

CANNON

®

Instrument Company recommends that laboratories check

their kinematic viscosity measurements with viscosity standards. If the
laboratory is using CANNON

®

calibrated viscometers and has developed

a good measuring technique, kinematic viscosity determination using a
standard will often point to temperature errors.

Viscosity standards should not be used to establish the correct tempera-
ture of the bath, however. Bath temperature should be checked and
corrected by applying the reliable thermometric techniques outlined
above.