Running an astm d2983 test, Table 8 — tools and materials required, Tesc operator's manual, rev 1.0 – Cannon Instrument TESC User Manual

Page 21

Advertising
background image

15

CANNON®

TESC Operator's Manual, Rev 1.0

Running an ASTM D2983 Test

This procedure runs sample conditioning programs and viscosity tests designed to meet
ASTM method D2983.

Table 8 — Tools and materials required

Description

Description

Assembled TESC System with mounted
Brookfield® DV2T viscometer

20 mL sample to be tested

25 mm × 150 mm rimless, disposable test
tube

Brookfield® #4B2 insulated spindle

Sample preparation materials

RS-485 to USB cable

USB flash drive with program files

1. Turn on DV2T and TESC. If the Watlow® controller starts to run a program, press the red

EZ

button to halt the test. (Steps symbol indicates a running program.)

Note:

For instructions on installing the Watlow® EZ-Zone Configurator

program, refer to page 20.

2. Load the appropriate thermal conditioning program into Watlow® controller:

a. Connect PC to Watlow® controller using RS-485 to USB cable.
b. Launch Watlow® configurator program:

Start Menu > EZ-ZONE Configurator > EZ-

ZONE Configurator

.

c. Select

Download a configuration file to a device

. Click

Next

.

d. Select the COM port to which the Watlow® controller is connected. Click

Next

.

e. Select

Load Program from File

.

f. Select program:

Advertising