Watlow® controller calibration, Install watlow® configurator program, Table 12 — brookfield® offset example – Cannon Instrument TESC User Manual

Page 26: Table 13 — tools and materials required

Advertising
background image

TESC Operator's Manual, Rev 1.0

20

Calibrating the TESC System

CANNON®

During testing, the Brookfield® viscometer measures and records the temperature of the TESC
base unit via an RTD probe located in the thermal core behind the test chamber. Adjusting
the offset of the Brookfield® viscometer only enables the recording of an accurate sample
temperature. It does not affect the operation of the TESC System.

Table 12 — Brookfield® offset example

Brookfield®
Display

Offset

Watlow® Display

Sample
Temperature

–41 °C

0

–40 °C

–40 °C

–40 °C

+1 °C

–40 °C

–40 °C

Watlow® controller calibration

The Watlow® controller must be calibrated before adjusting the temperature probe offset of
the Brookfield® viscometer.

Table 13 — Tools and materials required

Description

Description

Assembled TESC System with mounted
Brookfield® DV2T viscometer

20 mL of an appropriate reference sample

25 mm × 150 mm rimless, disposable test
tube

Brookfield® #4B2 insulated spindle

Sample preparation materials

RS-485 to USB cable

USB flash drive with program files

Install Watlow® configurator program

A copy of the Watlow® EZ-ZONE Configurator 5.0 program is provided with the TESC
software package. This program must be loaded onto a computer running a Windows®
operating system prior to calibrating or operating the TESC System.

1. To load the Watlow® EZ-ZONE Configurator 5.0 program, run

USB flash drive> Watlow >

EZCv5.exe

.

2. Follow the instructions provided by the installation wizard. For support and additional

information, visit http://www.watlow.com.