Related documents, Table 1 — tesc thermal conditioning programs, Table 2 — related documents – Cannon Instrument TESC User Manual

Page 8

Advertising
background image

2

Overview

CANNON®

TESC Operator's Manual, Rev 1.0

through the typical range of spindle speeds for the sample type or expected viscosity. This
eliminates the need to run multiple tubes of a sample. Once the viscosity measurements
finish, the TESC System returns the sample to room temperature, nominally 25 °C.
Afterwards, the operator can review the data and enhance the digital record with notes and
additional information. The TESC System comes with a set of thermal conditioning programs
for all of the common specification test temperatures as well as certain OEM measurement
and report requirements.

Table 1 — TESC Thermal conditioning programs

Fluid

Preheat Temperature

Test Temperature

Low viscosity (ATF, hydraulic fluid)

50 °C

–10 °C

–20 °C

–30 °C

–40 °C

High viscosity (gear oil)

90 °C

–12 °C

–26 °C

–40 °C

Related documents

Table 2 — Related documents

Title

Reference Number

Standard Test Method for Low-Temperature Viscosity of
Lubricants Measured by Brookfield® Viscometer

ASTM D2983

Automotive Gear Lubricant Viscosity Classification

SAE J306