4.6.3 External pulse output

The unit will also output a signal that is related to the flash rate of the unit. This signal can be used to control

additional stroboscopes so that each flashes at the same rate or to send a pulse to data collectors.

External circuit: Open collector output, Voltage below DC30V, Current below 30mA

Output pulse width: Approx. 200us

1pin

2pin

Inside DT-329

Photocoupler
(The same with TLP127
Toshiba made)

(36V)

[Output circuit]

[Output time]