5 Operation of the NucleoCounter

25

elevated signal levels. If the number of pixels is above a predefined limit then the
cassette insertion slit can be contaminated with an interfering particle, like a dust
particle. An error message is shown in the display and a cleaning of the insertion slit is
recommended (refer to chapter 7 Maintenance of NucleoCounter).

If a PC with NucleoView is connected to the instrument a contaminant is usually
observed as an object, which is stationary in all images.

To initiate a

Zero Count Check type in “F50 +

”. Then the following text will appear in

the display:

0 count check

0 count check

0 count check

0 count check

Continue

Continue

Continue

Continue

Esc Ab

Esc Ab

Esc Ab

Esc Abort

ort

ort

ort

Figure 20 Zero Count Check, step 1

Press “↵

” to continue. Now the instrument prompts you to remove any cassette from

the insertion slit:

Remove any

Remove any

Remove any

Remove any

cassette

cassette

cassette

cassette

Continue

Continue

Continue

Continue

Esc Abort

Esc Abort

Esc Abort

Esc Abort

Figure 21 Zero Count Check, step 2

Press “↵

” to continue. Now the instrument prompts you to close the lid in order to avoid

false light from entering the insertion slit during the Zero Count Check:

Close lid

Close lid

Close lid

Close lid

Continue

Continue

Continue

Continue

Esc Abort

Esc Abort

Esc Abort

Esc Abort

Figure 22 Zero Count Check, step 3

Press “↵

” to continue. Then a “Wait” message appears in the display:

wait...

wait...

wait...

wait...


Figure 23 Zero Count Check, step 3A