9 technical specifications – ChemoMetec YC-100 User Manual

Page 49

Advertising
background image

9 Technical specifications

37

9 Technical specifications

9.1

Technical specifications for NucleoCounter YC-100

Specificity

Cells stained with the DNA specific fluorescent dye, PI.

Sample types

Suspension of dispersed yeast cells.

Sample consumption

A minimum of 100µl is recommended for the NucleoCassette
loading. Approximately 50 µl of sample is loaded into the
cassette.

Analysis volume

Approximately 1 µl of the sample is being analyzed in the
NucleoCounter YC-100.

Measurement range

Measurement range:
Optimal Measurement range:

5x10

3

to 6x10

6

cells/ml

5x10

4

to 6x10

6

cells/ml

Presentation of Result

The NucleoCounter YC-100 presents the estimated cell density
in the sample in the cassette. In order to obtain the cell
concentration in the original cell suspension the result must be
multiplied with the appropriate multiplication factor to
compensate for any dilution.

Operation

Menu-controlled by means of keyboard and LCD display.

Analysis time

When pressing “Run

Run

Run

Run” on the NucleoCounter YC-100 the result

will be displayed within 30 seconds.

Physical data

Weight
Height
Width
Depth

3 kg
26 cm
38 cm
22 cm

Power

See Section 10.2 Power Supply

Input
Fuse

12VDC (11-13VDC)
UL Listed fuse:
600mA/250Vac (UL 248)

Power consumption
NucleoCounter

Peak
Ready mode
Standby

25 W
2.5 W
2 mW

Operation conditions

Maximum relative humidity 80 percent for temperatures up to
31°C decreasing linearly to 65 percent relative humidity at
maximum 35°C; minimum temperature 15°C.

USB

USB, version 1.1.
Note: Does not support USB Hubs.

Printer

Thermal or Impact printer with RS-232 interface and cable
supplied by ChemoMetec A/S.