5 Operation of the

991-0251 Rev. 1.0

Feb-2013

14

5.3

Result presentation

Depending on the type of analysis performed the result will be present ed on the PC within

a few seconds and the fluorescent image(s) will be stored on the hard disk. Typical process

time is from 20 seconds to 2 minutes for multi image and bi-color analysis.

5.4

Protection of the optical system

The lid covers the sample tray. This lid is designed to protect the area from dust and other

potential contaminants, and to keep external light from interfering with the fluorescent

image recording. The lid is spring loaded so it will automatically close during insertion of

the sample tray.

Unless inserting or ejecting a sample media or cleaning the sample tray, the sample tray

should be inserted at all times to protect the optical system of the NucleoCounter® NC-

250™. After operation, shut down the NucleoView™ NC-250™ software first, so the sample

tray will be inserted in case left ejected. Then the NucleoCounter® NC-250™ can be turned

off by removing the DC power plug or disconnecting the wall outlet.