8 Troubleshooting - Error messages

991-0251 Rev. 1.0

Feb-2013

20

Figure 13. Location, removal and inspection of the fuse

Disconnect the power supply from the NucleoCounter®. Then remove the fuse holder using

a screwdriver as indicated in Figure 13 (turn CCW). Inspect the fuse. If it needs

replacement use one as described by ChemoMetec A/S.

If the NucleoCounter® NC-250™ does not turn on after replacement of the fuse, or if the

fuse was not broken, please contact ChemoMetec A/S or the local distributor for further

instructions.