8 Troubleshooting - Error messages

991-0251 Rev. 1.0

Feb-2013

19

8 Troubleshooting - Error messages

Under certain conditions the NucleoCounter® NC-250™ will display error messages during

operation. Corrective actions are suggested below, but if they do not correct the errors,

contact ChemoMetec A/S or the local distributor.

Under certain conditions the NucleoView™ NC-250™ software will display error messages

during operation. Corrective actions can be found in the NucleoView™ NC-250™ software

Users Guide.

8.1

Required hard disk space for data storage.

The CM files stored on the hard disk can vary from 1 5MB up to in some cases more than

50MB. Therefore it is important from time to time to check whethe r enough free space is

present on the hard disk. ChemoMetec recommends a minimum of 10GBs free space on

the hard disk.

It is recommended to make a periodic archiving of CM files to external storage. Those

archived CM files can then be removed from the har d disk of the PC.

The periodic archiving can be made on either the creation date of the CM file, or when the

free disk space is below a certain amount, or when the disk space used for CM files

exceeds a certain amount. Contact the local IT Administrator in order to get information

about procedures for external data storing.

A good example of the archiving procedure could be based upon:

CM files older than 3 months are archived

CM files uses more than 50GBs of disk space, archive the oldest CM files until le ss

than 20GB is used

When less than 10 GBs free space, archive the oldest CM files until less than 20GB

is used.

Refer to the NucleoView™ NC-250™ Software User’s Guide

5

for additional information.

8.2

Power-on failure

If the NucleoCounter® NC-250™ cannot be turned on, please verify that the power supply

is connected to the NucleoCounter® NC-250™ and a working mains plug.

If the NucleoCounter® NC-250™ still does not turn on, inspect and/or replace the fuse as

described below.

The fuse holder is located at the rear side above the power plug, see Figure 13

5

Locate the file ”Install_Guide.html” on the USB Flash drive and follow the directions.