OPERATION AND INSTALLATION MANUAL

Page 11

Resistance to
Ground Test

Problem:Electrical leakage is creating an unstable
output from the instrument.

Cause:Water contamination in the load sensors or
cables.

Remedy:

Step 1.

Tie together the load sensor excitation (2),
signal (2) and ground (1) wires.

NOTE:

Be careful NOT to include the two C2 wires.

Step 2.

Use a megohmmeter and measure the resis-
tance between all five wires tied together
and the load cell metal body. The measured
value should be 5,000 megohms or more.

Step 3.

Replace the load cell if the cell fails both
tests.

WARNING

WHEN USING A MEGGER DO NOT EXCEED
50 VOLT RANGE.

If the sensor fails this test, remove the ground wire
and test with only the four live leads.

Replace the load cell if the cell fails both tests.

If the sensor passes the test, an insulation problem in
the cable is most likely.

Electrical Termination
Cable Color Codes

The cable is 6 conductor, shielded (floating) and 30
feet in length.

EXC+Red EXC -Black

SIG +Green SIG -White

C2+Gray C2-Violet

SHIELDYellow

Model Numbers

NOTE:

The -43 indicates a stainless steel Load Sensor with
stainless steel mounting hardware. For zinc plated
mounting hardware use -45.