Universal scale electrical connections, Universal scale electrical connections -3, Fig. 3-3 leveling scale -3 – Hardy ANY-WEIGH Bench Scales User Manual

Page 15

Advertising
background image

Chapter 3 - Installation

3-3

FIG. 3-3 LEVELING SCALE

Step 4.

Rotate the spirit level 90° and check for
level again.

Step 5.

Adjust the leveling feet if necessary.

Universal Scale
Electrical Connections

Make sure the cable is not pinched and is clear of the
feet. The Bench Scale requires a recommended DC or
AC excitation voltage of 5 volts with a maximum of
15 volts. Wire the Scale cable to an instrument or
indicator following the color code label found on the
upper frame or as follows:

+ Excitation

Red

+ Sense

Blue

+ Signal

Green

+ C2 Cal

Gray

- Signal

White

- Sense

Brown

- Excitation

Black

- C2 Cal

Violet

WARNING

USE CAUTION WHEN WIRING. MAKE SURE
TO DISCONNECT ALL POWER TO THE
SCALE. FAILURE TO REMOVE POWER CAN
RESULT IN SCALE OR INSTRUMENT DAM-

Spirit Level