Test weight calibration from rs networx, Test weight calibration from rs networx® -3 – Hardy ANY-WEIGH Bench Scales User Manual

Page 31

Advertising
background image

Chapter 6 - Calibration

6-3

Step 3.

Double click on the Node Icon of the
weigh module you want to calibrate. For
example: Node 36. The DeviceNet Weigh
Module Dialog Box appears.

Step 4.

Click on the Parameter tab. The Parameter
List appears with the information for the
weigh module at address 36.

Step 5.

Click on Cal LO.

Step 6.

Enter the Cal LO value.

Step 7.

Click on the Apply button.

Step 8.

Click on Yes.

Step 9.

Click on C2 Cal Cmd.

Step 10.

Set the C2 Cal Cmd to “1”.

Step 11.

Click on the Apply button.

Step 12.

Click on Yes.

Step 13.

Click on Cmd Save non-vol.

Step 14.

Set the Cmd Save non-vol to “1”.

Step 15.

Click on the Apply button.

Step 16.

Click on Yes.

Step 17.

The calibration is complete.

NOTE:

Requires C2

®

load sensors or load points.

NOTE:

RS NetWorx

®

is a registered trademark of Rockwell

Automation.

Test Weight
Calibration from RS
NetWorx

®

Step 1.

On the PC open RS NetWorx

®

Step 2.

Browse the Network.

Step 3.

Double click on the Node Icon of the
weigh module you want to calibrate. For
example: Node 36. The DeviceNet Weigh
Module Dialog Box appears.

Step 4.

Click on the Parameter tab. The Parameter
List appears with the information for the
weigh module at address 36.

Step 5.

Click on Span Weight.

Step 6.

Enter the Span Weight Value.

Step 7.

Click on Cal LO.

Step 8.

Enter the Cal LO weight.

Step 9.

Click on Cal LO command.

Step 10.

Set the Cal LO to “1”.

Step 11.

Click on the Apply button.

Step 12.

Click on Yes.