Head, Led inspection function, Chapter 5 environment settings – KEYENCE LS-7000 User Manual

Page 124: Press the

Advertising
background image

5-10

Chapter 5 Environment Settings

5

Head

Take the following steps to set the LED inspection function and mutual interference preven-
tion function.

Setting Item
• LED inspection function:

The intensity of the LED light source can be
inspected.

• Mutual interference prevention function: Prevents mutual interference that may result

when two measuring heads are located at
right angles.

Measuring Head Settings

1

Press the

FUNC

key twice when the program display is on the screen.

LED Inspection Function

Use this function to check for change of the LED light source or the fluctuation of light
intensity caused by dirt on the cover glass. It is recommended to make a regular check
on the change of light intensity after the installation of the LS-7000 Series.

Check on Intensity of LED Light Source
With the LED inspection function, the distribution of light will appear on the monitor as
shown below, which makes it possible to check the change of light intensity after the
LS-7000 Series is installed.

TIM

LO

GO

HI

OUT 2

OUT 1

FOCUS

2

1

A R E A

POSITION

HI

GO

LO

Indicates that 100%
light is distributed.

This bar indicates
5% intensity.

The distribution of light is uniform.

TIM

LO

GO

HI

OUT 2

OUT 1

FOCUS

2

1

A R E A

POSITION

HI

GO

LO

The display indicates that the light
intensity drops by 25%.

If the intensity of light does not recover with the cover glass cleaned or the mounting
position corrected, contact your Keyence representative.