Focus monitor, Setting method – KEYENCE LS-7000 User Manual

Page 66

Advertising
background image

4-14

Chapter 4 Function Settings

4

The input value is entered and the value in the next line flashes.

(3) Input the end edge (pitch) number.

4

Press the

ENT

key.

The input value is entered and the area display appears again.

Focus Monitor

The focus monitor is a function that makes it possible to check whether the target is lo-
cated within the range where the highest precision measurement of the target is ensured.
The focus monitor range is the distance between the receiver and target in the measuring
area. When the target is within the range specified, the FOCUS indicator will light.

FOCUS display is lit

The center of the measuring area

Focus display is OFF

Setting Range
The value is set at 10 levels between 0 and 9. Set the value to 0 if the focus monitor
function is not used. Refer to Focus Characteristics on page 8-6 for the set values and
focus range in detail.

Tips

The result of focus monitoring can be externally output as function output.

Refer to Selection of Function Output on page 4-17 for details.

Setting Method

1

Select the area to be set and press

the

ENT

key.

2

Press the

FUNC

key.

HI

GO

LO

PROG No.

PRINT

AREA

DISP

ESC

PROG

FUNC

ENT

H

Z

FOCUS

2

1

A R E A

POSITION

3

Press the

or

key and set the

value.

4

Press the

ENT

key after the value is

set.
The input value is entered and the area
display appears again.

Advertising