KEYENCE LS-3100/3100W Series User Manual

Page 58

Advertising
background image

52

CHAPTER 4 Functions and Controls

Entry Examples
• To specify "1" for program No., "Y" for segment selector, "1" for scanning

head No., and SEG (3, 4) for measurement segment: "P1Y1S3,4 CR"

• To specify "2" for program No., "X" for segment selector, and "OFF" for

segment check: "P2XSR CR"

• To specify "3" for program No., "X" for segment selector, "1" for scanning

head No., and "DIA" for measurement segment: "P3D CR"

• To set the number of measurements for averaging to 1,024, while not

changing the other settings on the front panel: "AA CR"

Setting Output Conditions

Sets the output conditions (processing methods for measured values) for the
desired program. The format differs with the setting item, as described
below.

Format:
( [P$] [C$] ) [COMMAND] CR

Note that one or more of the commands in ( ) can be omitted.
When [P$] is omitted, the current program No. will be valid.
When [C$] is omitted, the current channel No. will be valid.

Channel No.

Function

Command

Description

Setting tolerance

IS

Sets tolerance differentiation

differentiation output

output to ON

IR

Sets tolerance differentiation
output to OFF

Setting tolerance

T3

Sets 3-level differentiation

differentiation level

T7

Sets 7-level differentiation (3 or 7)

Setting tolerance

TAn

Sets value ("n") for HIGH3

range for 7-level

TBn

Sets value ("n") for HIGH2

differentiation

TCn

Sets value ("n") for HIGH1

TDn

Sets value ("n") for LOW1

TEn

Sets value ("n") for LOW2

TFn

Sets value ("n") for LOW3

Setting values

ASm, n

Sets "m" for +6 V, and "n" for -6 V

("m" and "n") for

(m > n)

analog output
voltage setting

[P$] [C$] List

[COMMAND] List

(To be continued)

Commands for setting functions

Function

Command

Description

Specifying program No.

P1

Specifies program No. 1

P2

Specifies program No. 2

P3

Specifies program No. 3

P4

Specifies program No. 4

P5

Specifies program No. 5

P6

Specifies program No. 6

P7

Specifies program No. 7

P8

Specifies program No. 8

P9

Specifies program No. 9

PA

Specifies program No. 10

Specifies output channel

V

Specifies output channel CH1

W

Specifies output channel CH2

Program No.

08.10.1, 10:56 AM

52

Advertising