KEYENCE LS-3100/3100W Series User Manual

Page 60

Advertising
background image

54

CHAPTER 4 Functions and Controls

Output of Measurement Data

Specifies the output method of the processed measurements which are
displayed on the main display (CH1) and sub-display (CH2), and outputs
them according to the specified method.

Input format: [COMMAND] CR

Function

Command

Description

Specifying data

X0

Outputs measurement data of CH1 once

output method

X1

Outputs measurement data of CH1

of data outputting

continuously

X2

Outputs measurement data of CH2 once

X3

Outputs measurement data of CH2
continuously

X4

Outputs measurement data of CH1 and
then that of CH2 once

X5

Outputs measurement data of CH1 and
CH2 alternately and continuously

X6

Stops continuous output

Output data format:
• At normal measurement
Measurement in "mm"

Measurement in "inch"

When there is no 1st, 2nd and 3rd data, signal corresponding to blank
( SPACE ) is output, allowing these digits to be displayed as blank.

• At erroneous measurement
Measurement in "mm" or "inch"

Measurement data processed on the basis of the current settings is output,
according to the specified command, 10 ms max. after the command has
been received.
When "X4" or "X5" command is entered, the values of CH1 and CH2 ob-
tained when the command is received by the LS are output in order of CH1
and CH2.

1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data

C

R

3

'-' or 10 digit

2

10 digit

1

10 digit

0

10 digit

-1

10 digit

Decimal point

-2

10 digit

-3

10 digit

-4

10 digit

1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data

C

R

2

'-' or 10 digit

1

10 digit

3

10 digit

-1

10 digit

-2

10 digit

Decimal point

-3

10 digit

-4

10 digit

-5

10 digit

1st data 2nd data 3rd data 4th data 5th data 6th data 7th data 8th data 9th data 10th data

C

R

-

-

-

-

-

-

-

-

-