2 resetting the volume calibration – A&D Weighing MPA-1200 User Manual

Page 32

Advertising
background image

30

[ 3 ] Press the Enter key to enter μL calibration

mode.


[ 4 ] Press the Up or Down key to alter the value

to the actual dispensed volume.

[ 5 ] Press the Enter key to confirm actual

dispensed volume.



[ 6 ] The notice mark lights up to indicate that

the volume calibration has been performed.
(In this example, the dispensed amount is
corrected and altered to 100μL.)

* After calibrating, a volume range that can be

selected may be limited depending on
available movement range of the piston.

11-2 Resetting the volume calibration

Go through the following procedure to restore the factory default settings for volume
calibration:

(The display is example.)

[ 1 ] Press the Back key.



[ 2 ] Press and hold the Back key for approx.

three seconds while “AUTO” is selected.




(μL calibration mode)