Changing the Display Brightness (10 adjustable brightness settings)

To change the brightness setting:

After the unit is powered on, depress and hold the power switch (for 2 seconds), the brightness will

then change by degrees. There are 10 levels of brightness.

Note: The default setting is level 4.

Lanyard Attachment

1. Thread thinner end of the lanyard through the hanging hole

at either side of the device.

2. Thread thicker end of the lanyard through the threaded end

before pulling it tightly.

Note: Unclasping the lanyard connector will facilitate thread-

ing thicker end through threaded loop.

Care and Maintenance

Replace the batteries promptly when low voltage indicator appears.

Clean surface of the fingertip and oximeter with 70% isopropyl alcohol before it is used in di-

agnosis of patients.

Remove the batteries if unit will not be operated for extended period of time.

It is best to store this product in a place where ambient temperature is -20°C -55°C (-4°F -

131°F) and humidity is < 93%.

It is recommended that the product should be kept in a dry environment at all times.

Please follow local ordinances when disposing of batteries.

Cleaning the Pulse Oximeter

Clean the rubber touching the finger inside of the Oximeter with a soft dampened cloth with 70%

isopropyl alcohol, and clean the test finger using alcohol before and after each test.

Note: Do not pour or spray any liquids onto the Oximeter, and do not allow any liquids to enter any

openings in the device. Allow the Oximeter to dry thoroughly before reusing.

Troubleshooting Possible Problems

Problem

SpO2% or pulse rate do not dis-
play normally.

SpO2% or pulse rate is shown
unstably.

The monitor cannot be powered
on.

Indication is
suddenly off.

Possible Reason

1. Finger is not inserted correctly.
2. Patient SpO2 value is too low to
be measured.

1. Finger might not be inserted
deep enough.
2. Excessive patient movement.

1. No battery or low battery power.
2. Battery not installed correctly.
3. The monitor may be damaged.

1. The Oximeter is automatically
powered off, when no signal was de-
tected after 8 seconds.
2. Battery power is too low to operate.

Solution

1. Retry inserting the finger.
2. There is excessive illumination.
3. Measure more times. If you determine
the product is working correctly, see your
healthcare provider for an exact diagnosis.

1. Retry inserting the finger.
2. Sit calmly and retry.

1. Replace battery.
2. Remove and reinstall battery.
3. Contact customer service center.

1. Normal.
2. Replace batteries.