American Diagnostic Corporation (ADC) Adimals 2150 User Manual
Page 6
“Err 3” or “Err 4”
“Err 7”
1. Low Power.
2. Recieving tube being shielded or dam-
aged together with broken connector.
3. Mechanical Misplace for recieve-
emission tube.
4. Amp Circuit malfunction
1. Low Power
2. Emission tube damaged
3. Current control circuit malfunction
1. Change batteries.
2. Return to Service Center
1. Change batteries.
2. Return to Service Center
Guidance and Manufacturer’s Declaration
Electromagnetic Emissions
Guidance and Manufacturer’s declaration - electromagnetic emission.
The Pulse Oximeter is intended for use in the electromagnetic environment specified below.
The customer or the user of the Pulse Oximeter should assure that it is used in such an environment.
Emission Test
RF Emissions CISPR11
RF Emissions CISPR11
Compliance Level
Group 1
Class B
Electronic Environment Guidance
The PULSE OXIMETER uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic equipment.
The PULSE OXIMETER is suitable for use in all establish-
ments,including domestic establishments and those directly
connected to the public low-voltage power supply, network
that supplies buildings used for domestic purposes.
Power consumption: Less than 40mA
Low power indication:
Battery Life: (2) ”AAA” 1.5V, 600mAh
alkaline batteries could be
continuously operated as
long as 30 hours.
Dimension:
Length: 1.9” (50mm)
Width: 1.1” (28mm)
Height: 1.1” (28mm)
Weight: 1.2 - 1.7oz. (35 - 50g)
(Including 2 AAA batteries)
Environment:
Operation Temperature: 41°F - 104°F (5°C - 40°C)
Storage Temperature: -4°F - 131°F (-20°C - 55°C)
Relative Humidty: 15%-80% in operation /
10%-80% in storage
Finger Range: .3” - .44” (7.64mm-11.19mm)
Minimum Weight: 33lbs. (15kg)
Technical Specifications
Display Type: OLED Display
SPO2 display range: 70-100%
PR Display Range: 30-235 BPM
PR Display Mode: Bargraph
Data Update Time: < 15 s
LED Wavelengths
Red: 660nm
Infrared: 940nm
Measurement range:
SPO2: 70-100% - 100% ±3%; s
≤70%
no definition
Pulse Rate: 30-235 BPM, ±2 bpm during
the pulse rate range of 30-99
bpm and ±2% during the pulse
rate range of 100-235 bpm.
Accuracy: ±2% on the stage of 80%-100%
±3% on the stage of 70%-79%
Accuracy: ±2 BPM or ±2% (larger)
Pulse Intensity: Bargraph Indicator
Power Requirements: Two AAA alkaline Batteries
Symbol Definitions
Important Warning/Caution
Latex-Free
Equipment type is BF
Hemoglobin Saturation
Heart Rate (BPM)
Low Power Indication