American Diagnostic Corporation (ADC) Diagnostix 3.5v User Manual

Page 11

Advertising
background image

Electrostatic discharge

(ESD) according to

IEC61000-4-2

± 6 kV contact
discharge

± 8 kV air discharge

± 6 kV contact

discharge
± 8 kV air

discharge

Immunity

Tests

IEC 60601

Test Level

Electromagnetic

Environment Guidelines

Guidelines and manufacturer's declaration - electromagnetic emissions

The Diagnostix

TM

instrument is intended for operation in an electromagnetic environment as specified below.

The customer or the user of the Diagnostix

TM

instrument should ensure that it is used in such an environment.

Note - UT is the alternating supply voltage prior to application of the test level.

Compliance

Level

Floors should be made of wood or
concrete or be covered with
ceramic tiles. If the floor is covered
with a synthetic material, the
relative air humidity must be at
least 30%.

Fast transient electrical

interference/bursts

according to

IEC61000-4-4

± 2 kV for power

lines
± 1 kV for input and

output lines

3 A/m

3 A/m

Not applicable

The quality of the supply voltage
should correspond to that of a
typical business or hospital
environment.

Surges

IEC61000-4-5

± 1 kV voltage

phase-to-phase
± 2 kV voltage

phase-to-earth

Not applicable

The quality of the supply voltage
should correspond to that of a
typical business or hospital
environment.

Voltage dips, short-time

interruptions and

fluctuations in the

supply voltage according

to IEC61000-4-11

Magnetic field at the

mains frequency (50Hz)

according to IEC61000-4-8

<5% UT (>95 % drop
in UT) for 0.5 cycles

40% UT (60 % drop
in UT) for 5 cycles

70 % UT (30 % drop
in UT) for 25 cycles

<5% UT (>95 % drop
in UT) for 5 s

Not applicable

The quality of the supply voltage
should correspond to that of a
typical business or hospital
environment.

If image disturbances occur, the
Diagnostix

TM

instrument may have

to be placed further away from the
sources of mains-frequency
magnetic fields, or magnetic
shielding may have to be installed:
the mains-frequency magnetic
field should be measured at the
intended set-up site in order to
ensure that it is small enough.