American Diagnostic Corporation (ADC) Diagnostix 3.5v User Manual

Page 9

Advertising
background image

the instrument head. Pull the lamp out of the instrument head with thumb and
forefinger or a suitable tool. Insert a new lamp.

CAUTION: The pin on the ophthalmoscope lamp must be inserted into the guide
groove on the instrument head.

*Use only ADC

®

or Riester lamps.

11. SPARE PARTS AND ACCESSORIES

For a complete list of our physical exam instrument spare parts and accessories,
please visit our website at www.adctoday.com.

12. CLEANING AND DISINFECTION

Cleaning and disinfection of the medical devices serves to protect the patient, the
user and third parties and to preserve the value of the medical devices. Due to the
product design and the materials used, no defined limit can be specified for the
maximum number of reprocessing cycles that can be carried out. The life span of the
medical devices is determined by their function and by gentle handling of the
devices. Defective products must undergo the reprocessing procedure described
before being returned for repair.

Cleaning and Disinfection
The instrument heads and handles can be cleaned externally with a moist cloth until
visually clean. Wipe disinfection as specified by the disinfectant manufacturer. Only
disinfectants with proven efficacy should be used, taking into account the national
requirements. After disinfection, wipe the instrument down with a moist cloth to
remove possible disinfectant residues. The components that come into contact with
the skin can be rubbed down with alcohol or a suitable disinfectant.

Please Note!
• Never immerse the instrument heads and handles in liquids!

Take care to ensure that no liquids get inside the casing!

• This item is not approved for automated reprocessing and sterilization.

Single-Use Ear Specula

For single Use Only

WARNING: Repeated use can cause infections.

13. TECHNICAL SPECIFICATIONS

Ambient Temperature:

32°F-104°F (0°C to +40°C)

Relative Humidity:

30% to 70% non-condensing

Transport and Storage

Temperature: 14°F-131°F (-10°C to +55°C)

Relative Humidity:

10% to 95% non-condensing